Abstract

Background: Percutaneous transcatheter closure (PTCC) of atrial septal defect (ASD) may convert to a long procedure. We aimed to identify predictors of prolonged procedure.

Methods: Under transesophageal echocardiography and fluoroscopy guidance, 81 children with ASD underwent PTCC. Retrospectively, medical charts, echocardiographic recordings, catheterization reports and fluoroscopic films were reviewed. Demographics, echocardiographic measurements of ASD, dimensions of the device and hemodynamic data were collected. Prolonged procedure was defined as the duration from device deployment out of the delivery sheath to its release exceeding 10 minutes. A statistical model was designed using stepwise logistic regression analysis. Receiver operating characteristic curves were plotted to find the best cut-off for significant predictors.

Results: The procedure was prolonged in 25 patients. By monovariate analysis, the significant predictors for prolonged procedure were smaller, and younger patients, larger ASD, smaller left atrial (LA) dimensions and device waist ratios to weight, patient’s length, and LA dimensions. By multivariate analysis, the significant predictors were deficient septal rim toward superior vena cava (SVC) and device waist diameter in relation to patient’s length (best cut-off: < 12 mm and > 0.13, respectively). In three cases (3.7%) the device embolized; retrospectively possibly the cause is small used device and in one combined with vigorous physical activity.

Conclusions: A short septal rim toward superior vena cava and large device waist size in relation to patient size and/or LA dimensions may predict prolonged procedure during PTCC of ASD.

PDF

Download the article PDF (381.54 KB)

Download the full issue PDF (21.51 MB)

Mobile-ready Flipbook

View the full issue as a flipbook (Desktop and Mobile-ready)

Cite this article as: El-Segaier M, Jadoon S, Javid T, Wani TA, Galal MO. Predictors of Procedure Time Prolongation During Percutaneous Transcatheter Closure of Atrial Septal Defect – A Retrospective Study. Structural Heart Disease 2018;4(3):69-78. DOI: 10.12945/j.jshd.2018.039.17

You must be registered and logged in to leave comments.

There have been no comments posted yet

All comments will be screened and reviewed before posting. Statements, opinions, and results of studies published in Journal of Structural Heart Disease are those of the authors and do not reflect the policy or position of The Journal and Science International and the Editorial Board and provides no warranty as to their accuracy or reliability. Material is copyrighted and owned by Science International and cannot be used without expressed permission.